GEOPIG Gas Chromatography Lab
The Varian CP-3800 Gas Chromatograph is setup for running gases and for
running liquid samples for dissolved organic analysis. The system has two
injectors, one for gases and one for liquids.
The gas analysis setup consists of a two packed
columns (Molsieve 13X80/100 and Poropak Q 80/100), a methanizer, a thermal
conductivity detector (TCD), and a flame ionization detector (FID). The
way it is plumbed now allows you analyze for hydrogen, carbon dioxide,
carbon monoxide, and methane in one run using the TCD detector and one of
the FID detectors. This setup can also be used to analyze for alkanes and
alkenes, as well as oxygen and other gases depending on you choice of
carrier gas (nitrogen or argon).
The dissolved organic analysis is done through a CP1177 split/splitless
injector, a Supelcowax-10 capillary column and one of the FID detectors.
With the column installed it is currently setup for looking at polar
organics such as alcohols, although it works for ketones and aldehydes.
Carrier gas is helium.
For both gases and liquids, the injectors are set up for manual
injection (i.e. there is no autosampler or injection valve).